Figure 6-6 Software Screenshot
6.8.6
Auto Cool Time
After a test is completed and auto cooling has been selected in the test screen (see
Start Test section), the cooling valve automatically opens and begins to cool the
cylinder and oil reservoir. It is possible to select how long a period of cooling is
active. Selections include between 1-24 hours and always on. If a test is started
while cooling is still activated, the test will not heat even if it is programmed to
heat. In other words, cooling must be deactivated if a test is to be started, even if it
is still in automatic mode.
6.8.7
Sample Rate
This button allows the user to select a sampling rate for taking data. Rate is given in
number of seconds between samples. To save data space and achieve acceptable
test resolution, a sample rate of 30 or 60 seconds is recommended. For increased
resolution lower the sample rate.
6.8.8
Temperature and Pressure Units
These buttons allow the user to select English or SI.