6.7 The Main Menu
Figure 6-1 Software Screenshot
The main menu is the starting point for controlling the instrument. From here users
may start new tests, setup test parameters, setup instrument parameters, or view an
old test. Also located on the main menu is the reading for current temperature,
consistency, and pressure. On the right side of the screen you will also notice the
version software that is currently installed on your instrument.
The cooling button is located on the main screen and is operated by selecting it to
turn it on. Select again to turn off. The cooling button is connected to a solenoid
valve which is connected to the cooling input located on the back of the instrument.
Typically, water is used at house pressure and temperature. However, you may also
connect a circulating chiller. If cooling is turned on when start test is activated, the
cooling will remain on and heating will not ensue. Turn cooling off if you wish to
start a normal test.
The motor switch is located on the main screen and is operated by selecting it to
turn it on. Select again to turn off. The motor speed is set to turn at 150 rpm.