When batteries need to be exchanged, do always exchange a set of two
batteries, exchange of a single battery causes limitation to operating time of a
table and quick wear of a new battery.
•
Hand support should be fixed on table strip with a multi-position grip which
makes angular movement of the support possible.
•
If it is necessary to replace oil or a battery (dangerous waste), one should follow
existing environmental protection regulations.
•
Hand controller is the obligatory accessories for surgical tables.
•
Hand controller must be within reach of the medical staff.
Notes concerning disinfecting
•
The product must not be disinfected in disinfecting chambers!
•
No bleaching agents (containing active chlorine or oxygen), caustic or corrosive
chemicals are allowed!
•
No agents destroying the structure of plastic (organic solvents) can be applied
to the plastic elements!
•
No disinfecting agents containing alcohol are allowed for cleaning of
polyurethane elements (mattresses).
•
Before disinfecting disconnect the table from mains pulling out the cable!
Disregarding the above requirements concerning cleaning and disinfecting shall
result in losing the guarantee for the product!