![Famed SU-14 User Manual Download Page 21](http://html.mh-extra.com/html/famed/su-14/su-14_user-manual_541744021.webp)
User manual for Operating table SU-14
Page 21
5 Criteria on whose basis it is assessed whether product
operation is correct or not
Caution!
Correctness of table operation shall be checked every day before the beginning of
its operation.
The method of checking whether operation of the table is correct:
1. Check stability of the table when its feet are down by trying to move the table manually ( when the
wheels are blocking): pushing of the table in any direction should not cause tilting or moving.
2. Check mechanisms of adjustment of positions of the segments controlled by gas springs by a
change of position and locking (see item 3 Operation). The mechanism should work without
jamming, after locking the segment should not change its position (check by manual pressing of
the segment).
3. Check if there are no mechanical plays by manually moving the top of the table.
4. Check the work of the electro-mechanics system by performance of all movements controlled by
wire-connected control.
If the table underwent the above described tests with positive results and there were no disturbing
sounds (squeaks and grinds), the table can be used safety.
Otherwise see item 6.6 ‘Location and removal of defects’.
Caution!
If the product is not fully functional, i.e. the output parameters differ from the
description contained in this manual, the bed must not be used. This situation
should be reported to the producer or supplier (dealer). The use of an improperly
functioning product may result in damages, which will encumber the user and
which will not be a matter of producer’s responsibility.