![Famed SU-14 User Manual Download Page 24](http://html.mh-extra.com/html/famed/su-14/su-14_user-manual_541744024.webp)
User manual for Operating table SU-14
Page 24
When to do the inspection:
– the device has been repaired;
– periodically (every 12 months) during technical inspection.
To ensure safe and correct operation of the device during the service life, the user must perform
technical checks listed below every 6 months.
Scope of technical condition check
- Check of all mechanical functions by executing all movements in line with the technical
specification parameters:
– Checking the table movement capabilities.
– Checking central wheel lock.
– Checking the removal and installation of removable segments.
– Checking the load bearing plates under mattresses.
– Check all visible bolted connections without removing housing.
– Check for mechanical play and if all locks work.
– Check the condition of the power supply cable.
– Check the insulation of wired controller cable.
– Check the condition of columns cover.
– Check condition of mattresses.
– Check efficiency of the central wheel lock.
An inspection should involve a visual inspection and the noticed malfunctions should be handled
as provided in item 6.6 Location of defects and their removal.
6.6
Location of defects and their removal
damage
Possible cause
removal
The table cannot be moved
locked wheels
Unlock blockade pedal
Batteries discharged
Charge the batteries
Control plug incorrectly
plugged to the socket in
the table
Plug in the plug correctly
Damaged pilot
Contact your service
Blockage of the driver
Disconnect the table of the alimentation,
wait few seconds and reconnect.
Plays in screw joints
Plays appeared during
product operation
Screw down with generally available
wrenches
Lack of possibility of using
the preset position
Change of position setting Insertion again the position to memory
according to point 3.5
If a fault can't be eliminated, abstain from using the product and report the matter in the Service
Department, or service of FAMED
Ż
YWIEC company.
6.7
Product liquidation
If the customer resigns from further product exploitation, he is obliged to product liquidation
according to rules of environment protection; detailed information is situated in annex no. 3.