User manual for Operating table SU-14
Page 19
to turn the table. All castors brake causes a total bed immobilization.
Caution!
Throughout surgical procedures table wheels should be blocked.
Table should be move by 2 people at least.
Do not move table with workload (patient) through inclines and outside of
buildings).
Table should be move in minimal height.
The floor under the table must be free from any obstacles!
When rolling the table avoid collisions!
Do not roll the table over electric cables!
3.13
Antistatic properties
A table structure allows for drainage of electrostatic charge through the following channels:
–
By antistatic wheels to conducting floor,
–
By potential equalisation clamp.
Operating table SU-14 should be used on antistatic floor. In case of lack of such floor, an electrostatic
charge is drained through a potential equalization line. The potential equalising conductor is a
standard table accessory.
Antistatic properties of the table shall be maintained if mattresses produced by FAMED Sp. z o.o. are
used.
3.14
Potential equalising clamp
Operating table have inside system of potential equalization leading to the clamp (fig. 1, item pos. 16)
marked by symbol
.
Caution!
Throughout surgical procedures the table must be connected to the installation of
potential equalization in the operations room.
The cable is provided in standard table accessories
3.15
Collisions
In some extreme positions of the table, in particular, when accessories installed on side stripes are
used, mechanical collisions may take place. Because of that, one should protect the table and
accessories from damages.
3.16
Battery charging
A charger built-in in the table makes it possible to charge the batteries in the table.
In order to charge batteries, connect a power supply cord to a socket in a table and set a switch (fig.1,
item 17) in position „I”. The batteries are charge automatically.
Caution!
If after 12 hours of charging and after disconnecting a power supply cord, it is
impossible to activate any electrical movements, this means that the table is
damaged. This fact should be notified to an authorized service centre.
Table work time is about 10 surgeries from full batteries.
It is required to charge batteries when there is a lack of possibility of carrying out movements or in the
case of noticeable slowdawn of movements.
Battery life time is 3 years.
Caution!
The table may not be stored with batteries low.
If batteries need to be exchanged, do always exchange a set of two batteries.
Exchanging a single battery limits an operating time of a table and causes quick
wear of a new battery.