![Famed SU-14 User Manual Download Page 4](http://html.mh-extra.com/html/famed/su-14/su-14_user-manual_541744004.webp)
User manual for Operating table SU-14
Page
4
•
Lock operation table in given position before transporting a patient.
•
When using the table close to medical equipment working on high frequencies
and defibrillators one should closely follow operating instruction for that
equipment. Improper operation may become a source of dangerous accidents.
There is a danger of serious burning of the patient through the contact with
metal parts of the table or its equipment.
•
Throughout surgical procedures the table must be connected to the installation
of potential equalization in the operations room.
•
You must make sure, that accessories are properly fixed to the tabletop.
•
When changing operation table's position you must avoid any collisions
between the table and accessories.
•
Not original accessories may be applied only after consulting with Famed
Ż
ywiec company about their use.
•
Operation table's electrical conductivity must be checked once a year. The
check should be carried out by trained service personnel.
•
Avoid endangering patient's respiratory system, neural pathways and
circulatory system by ensuring the patient is in proper position and by
observing his/her condition.
•
When performing longitudinal (Trendelenburg or reverse Trendelenburg) and
lateral tilt, patient should be secured against uncontrolled sliding down from the
operating table. Use shoulder supports, supporting rollers, side supports, belts
and grips as securing elements.
Notes concerning: start-up, operation and use
•
The table, during use, should not be in the place which obstructs its
disconnection from the mains!
•
When performing longitudinal (Trendelenburg or reverse Trendelenburg),
backrest should be in zero position or above the horizontal level.
•
When anti-trendelenburg function is performed, footrests may collide with the
basis.
•
When performing Trendelenburg function or functions downward movement
may take place collision with the base of the backrest ( backrest tilted down).
•
Adjusting angle position of a back rest requires manual operation.
•
The table should be moved by 2 people at least.
•
Do not move table with workload (patient).
•
The table should be moved in minimal height.
•
The floor under the table should be free from any obstacles.
•
When rolling the table avoid collisions.
•
Do not roll the table over electric cables.
•
The table has to be connected to mains in consistence with its name plate.
•
Do not use the cable when it is damaged or its insulation is worn out.
•
Do not connect the table to mains in places where there is a danger of an
explosion.
•
Do not store a table with flat batteries (a red diode is illuminated on a controlling
device during any movement).