In tables: item 7 Characteristics of electromagnetic environment – electromagnetic environment in
which medical device the table SU-14 should be used. Recommendations and warnings which should
be followed by the users were also presented.
Caution!
Use of different accessories, additional equipment, cables, spare parts than those
offered and/ or recommended by the producer may cause an increase of emission
and/ or decrease of bed’s resistance to all electromagnetic phenomena.
Recommended distances between portable radio-transmitters and the product
Rated maximum
output power of
transmitter
W
150 kHz to 80 MHz
P
d
2
,
1
=
distance in meters
150 kHz to 800 MHz
P
d
2
,
1
=
distance in meters
800 MHz to 2.5 GHz
P
d
3
,
2
=
distance in meters
0.01
0.1
0.1
0.2
0.1
0.4
0.4
0.7
1
1.2
1.2
2.3
10
4
4
7
100
12
12
23
For transmitters, the maximum output power of which is not specified above, the separation
distance should be calculated according to the formulas provided. P is a power in watts (W)
according to the declaration of the transmitter manufacturer.
NOTE The above guidelines may not be applicable to all cases. Propagated electromagnetic
waves are absorbed and reflected from buildings, objects and people.