8.0 TROUBLESHOOTING
This TRACHEOFIRST
®
COMPACT device has been subjected to a thorough inspection at the factory. Should a malfunction
nevertheless arise, you may possibly be able to rectify it yourself if you observe the following instructions.
Symptom of malfunction
Possible cause
Remedy
Device fails to start
operating
Mains plug of the power supply is not well
seated
Check all plug connections.
Vacuum in the container
De-aerate container
Insufficient power
Filter is obstructed
Replace filter
Leak in the tubing or the secretion container
lid
Check secretion container lid and tubing for
firm seating
1. Weak vacuum or no
vacuum is indicated
1.1 Bacterial filter is missing
Insert bacterial filter
1.2 Leak in the tubing or on the secretion
container lid
Check connections of connecting tube/ bac-
terial filter
Check assembly position of the secretion
container lid
Check secretion container lid and tubing for
firm seating
1.3 Secretions or blood have been aspirated
into the device and the valve platelets of the
unit are sticking
If this is the case, the device must be sent in
for repair
2. Strong vacuum is
indicated
2.1 Bacterial filter is obstructed
Replace bacterial filter
2.2 The float of the overflow protector is blo
-
cking the secretion container lid inlet
Check inlet of secretion container lid. If
necessary, empty container, clean overflow
protector and check float for free movement