![F.W. Bell 8000 series Instruction Manual Download Page 87](http://html.mh-extra.com/html/f-w-bell/8000-series/8000-series_instruction-manual_535141087.webp)
F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 8 – Remote Operation
8-8
Functional
Description
This remote display is separated into a series of different windows that
display meter information including common controls, the flux density,
frequency, and temperature measurements, data logging, and file utilities
for the internal flash drive.
Each window can be collapsed or expanded to display the information of
interest. This is done by clicking the double arrow symbols in each
window’s header.
Settings
The Settings window, allows the remote user to set the meter’s Flux
Units, Angle Units, Time Units, Temperature Units, and the Display
Update Rate.
NOTE: The update rate is the meters display update rate and is also used as
the data log rate. It is NOT the update rate for this application, the update rate of
this application is limited to twice per second and may vary based on network
speed and or available bandwidth. In either event this will not affect the meters
display or data log update rates.
Date and Time
The Date and Time window, allows the remote user to manually enter
and set the meter’s date and time, or use the time from the remote
computer to set the meter’s date and time.
Meter Info
The Meter Info window shows the meter’s model number and installed
firmware versions.
Data Logging
The Data Logging window can start, stop, or cancel a data logging
session to the meters internal flash drive.
Pressing <Start> will start the data log and show the number of rows
captured, note that the rate at which the rows are captured is set by the
Update Rate setting.
Pressing <Stop> will terminate the open log session, save the file on the
internal flash drive, and open a pop up window that allows the user to
open or save (download) the log file to the remote computer.
Pressing <Cancel> will terminate the log session and discard the log file.