Section 8 – Remote Operation
8-36
value could have been previously defined using the “:INPut#:OFFSet
<NRf>” command, manually entered using the RELATIVE menu, or
generated automatically using the probe as the input source.
When <b> is “ONCe” the automatic RELATIVE operation is started. This
command initiates the same process as the front-panel RELATIVE push
button. The Hall probe should be placed in a stable environment prior to
issuing this command. The value presently measured by the probe will
become the new relative offset value. This will overwrite any value
previously defined using the “:INPut#:OFFSet <NRf>” command or entered
manually from the RELATIVE menu. When the automatic RELATIVE
operation is complete the RELATIVE function will be enabled.
:INPut#:OFFSet:STATe?
Returns the state of the RELATIVE function “OFF” or “ON”.
Analog Output
Commands
The instrument provides a corrected and uncorrected analog output for
each channel. The 8030 models also provide an analog output for the
vector summation. These commands specify the output configuration by
setting the output mode and scaling options.
Note: The Vector Sum output does not have a waveform output mode, and
can only be turned on and off. Selecting waveform or RMS modes will turn
it on.
Since these commands are channel-specific the “#” in each command
should be replaced with the channel number 1, 2, or 3. If the channel
number is omitted then Channel-1 is assumed or controls the vector
summation output.
:OUTPut#:UNCorrected:MODe <n>
:OUTPut#:CORRected:MODe <n>
Sets the mode of operation for the selected uncorrected or corrected
analog output where:
n = 0 >> Output Off
n = 1 >> Output is in Waveform Output Mode
n = 2 >> Output is in RMS Output Mode
:OUTPut#:UNCorrected:MODe?
:OUTPut#:CORRected:MODe?
Returns the mode of operation for the selected uncorrected or corrected
analog output where the return value represents the following:
0 = Output Off
1 = Output is in Waveform Output Mode
2 = Output is in RMS Output Mode