F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 8 – Remote Operation
8-31
:SENSe#:FLUX:DC
Specifies that the measured flux density mode will be static (DC) or
alternating (AC). These commands do not alter the present range setting.
:SENSe#:FLUX:RANGe:FIXed <n>
:SENSe#:FLUX:DC:RANGe:FIXed <n>
:SENSe#:FLUX:AC:RANGe:FIXed <n>
Specifies that the measured flux density mode will be static (DC) or
alternating (AC) while also specifying one of four fixed operating ranges for
the instrument, as defined in table shown below.
Table 8-17 Operating Ranges with Various Probes
<n>
Maximum Flux Density
Using Low Field Probe
Maximum Flux Density
Using Mid Field (1x) Probe
Maximum Flux Density
Using High Field (10x) Probe
0
300 mG (30
µ
T)
30 G (3 mT)
300 G (30 mT)
1
3 G (300
µ
T)
300 G (30 mT)
3 kG (300 mT)
2
30 G (3 mT)
3 kG (300 mT)
30 kG (3 T)
3
300 G (30 mT)
30 kG (3 T)
300 kG (30 T)
Range and Mode
Commands
:SENSe#:FLUX:RANGe:AUTo <b>
:SENSe#:FLUX:DC:RANGe:AUTo <b>
:SENSe#:FLUX:AC:RANGe:AUTo <b>
Specifies that the measured flux density will be static (DC) or alternating
(AC) and that AUTO ranging is turned OFF when <b> is “0” or ON when
<b> is “1”. When AUTO range is turned on the instrument determines the
range that will offer the best resolution. When AUTO range is turned off
the range will be fixed to the last range that was in use when the command
was issued.
:SENSe#:FLUX:RANGe?
Returns an ASCII string representing the most present or most recent
AC/DC mode setting, the present range, and the state of the AUTO range
feature.
The string will start with the phrase “AC” or “DC, followed by a comma(,).
The next single digit will be the range setting, as shown in Table 8-18
,
and
will be in the range of 0 to 3, followed by a comma(,). Next will be the
phrase “OFF” or “ON, indicating the state of the AUTO range feature.
For example the string DC,2,OFF indicates that the channel is in DC mode
on a fixed range setting of 2 or 30G for a Low Field Probe, 3kG for a 1x
probe etc.
Display Update
Command
This command controls the display update rate as well as the software
averaging used by the instrument.