F.W. BELL 8000 Series Gauss/Tesla Meter Instruction Manual
Section 6 – Flux Density Measurement
6-4
Measurement
Range
Selection
Auto Range
In the Auto Range mode, the range is advanced if the reading reaches 90%
full scale of the present range. The range is lowered if the present reading
falls below 8% full scale of the present range.
NOTE: When the word “
Auto
” does not appear on the Measurement
Range indicator line, the channel is in manual ranging mode.
Over Range Condition
An over range condition will occur when the following is true:
The instrument is in manual range mode or is in auto range mode
at the highest range; and the measured flux density exceeds 110% of
the present full scale range.
When an over range condition occurs the instrument will display the
message “
Overrange”
in place of the flux density reading.
The next
higher range should be selected. If already on the highest range, then the
flux density is too great to be measured with this instrument and its probe.
Note: For dc mode operation, the polarity (+/-) of the flux density that
caused the over range condition will be displayed to the left of the
“
OverRange”
message.
Note: Using Auto Range with high crest factor signals or pulses may cause
toggling of the range setting if the peak is considerably higher than the
average or rms value of the signal and exceeds the full scale value. For this
type of signal, it is best to use a fixed range setting. In firmware v1.6.0 and
newer, if the signal is very large the
CLIP
indicator may appear on the
display indicating a signal which may have a low average or rms value, but
is overdriving the amplifier. If this indicator appears, you should probably
use the next higher range.
Measurement
Mode
Selections
The 8000 Series Gaussmeter is capable of measuring either static (DC) or
alternating (AC) magnetic fields. To measure DC fields the meter must be
set to operate in DC mode and conversely to measure AC fields the meter
must be set to AC mode.
The active measurement mode is displayed in the upper right-hand corner
of the display and indicates that the meter is in either DC or AC
measurement modes.
To change the measurement mode:
1) Using the channel keys, press the [Mode] button for the channel you
want to change.
2) Navigate the menu system and select the channel and mode
desired.