![F.W. Bell 7030 Instruction Manual Download Page 97](http://html.mh-extra.com/html/f-w-bell/7030/7030_instruction-manual_535140097.webp)
F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
Section 7 – Remote Operation
7-23
SCPI COMMANDS-
ERROR QUEUE
MESSAGES AND
COMMANDS
As error messages occur, they are placed in the error queue. Each
message will contain a number, a comma (,) and a brief description of the
error. Negative (-) numbers are used for SCPI defined messages while
positive (+) numbers relate specifically to the instrument.
The error queue can hold up to 10 messages. Each time the queue is
read the oldest message is presented and removed from the queue. If
no errors exist the message “0, No error” will be returned. If the queue is
full the message “-350, Queue Overflow” will occupy the last queue
location. It is an indication that at least one, perhaps more error
message(s) were lost.
There are certain error status bits that will set in the STANDARD EVENT
register. These bits provide general error indications. The error queue
will provide more detailed information about the errors.
:SYSTem:ERRor?
:SYSTem:ERRor:NEXT?
This command places the oldest error message in the output queue and
removes it from the error queue.
SCPI COMMANDS–
SYSTEM
INFORMATION
AND
CONFIGURATION
COMMANDS
:SYSTem:VERSion?
This command places the version number of the SCPI standard that is
supported by this instrument in the output queue. The version number is
in the format of “YYYY.V”, where “YYYY” is the year version of the
standard and “V “is the approved revision number, such as “1999.0”.
:SYSTem:TIMe <NRf>, <NRf>,<NRf>
The instrument’s real time clock can be set from the DISPLAY menu or
with this command using the following format:
:SYSTem:TIMe <hour>, <minute>, <second>
where:
<hour> is in the range of 0 – 23
<minute> is in the range of 1 – 59
<second> is in the range of 1 – 59
Summary of Contents for 7030
Page 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Page 24: ......
Page 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Page 26: ......
Page 122: ...THIS SIDE BLANK Back B 3 ...
Page 124: ...THIS SIDE BLANK Back of Warranty ...
Page 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Page 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...