Section 2 - Specifications
2-1
Section 2
Specifications
INSTRUMENT
Full-scale ranges are shown in the
Tables 2-1a – 2-1c
below; listed by probe type. In all cases, the
resolution is 1 part in 300,000.
Table 2-1a: Ranges for Low Field Probe:
gauss (G)
tesla (T)
oersted (Oe)
ampere-turn/meter (A/m)
300.000 mG
30.0000
µ
T
300.000 mOe
23.8732 A/m
3.00000 G
300.000
µ
T
3.00000 Oe
238.732 A/m
Note: Low Field probe cannot be used for measurements above 2 Gauss.
Table 2-1b: Ranges for Mid Field Probe:
gauss (G)
tesla (T)
oersted (Oe)
ampere-turn/meter (A/m)
30.0000 G
3.00000 mT
30.0000 Oe
2.38732 kA/m
300.000 G
30.0000 mT
300.000 Oe
23.8732 kA/m
3.00000 kG
300.000 mT
3.00000 kOe
238.732 kA/m
30.0000 kG
3.00000 T
30.0000 kOe
2.38732 MA/m
Table 2-1c: Ranges for High Field Probe:
gauss (G)
tesla (T)
oersted (Oe)
ampere-turn/meter (A/m)
300.000 G
30.0000 mT
300.000 Oe
23.8732 kA/m
3.00000 kG
300.000 mT
3.00000 kOe
238.732 kA/m
30.0000 kG
3.00000 T
30.0000 kOe
2.38732 MA/m
300.000 kG
30.0000 T
300.000 kOe
23.8732 MA/m
Summary of Contents for 7030
Page 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Page 24: ......
Page 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Page 26: ......
Page 122: ...THIS SIDE BLANK Back B 3 ...
Page 124: ...THIS SIDE BLANK Back of Warranty ...
Page 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Page 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...