F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
7-38
Section 7 – Remote Operation
SCPI COMMANDS-
HOLD COMMANDS
(Continued)
:CALCulate#:HOLD:MAXimum:STATe <b>
The MAX HOLD function is turned on when <b> is “1” or “ON”, or turned off
when <b> is “0“ or “OFF”.
:CALCulate#:HOLD:MAXimum:STATe?
The state of the MAX HOLD function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate#:HOLD:MAXimum:CLEar
The value being held by the MAX HOLD function is reset to 0.
:CALCulate#:HOLD:MAXimum?
The value being held by the MAX HOLD function is placed in the output
queue and will be between 0.0000000000 and
±
29999999. The meaning
of this value depends upon the currently selected unit of measure (gauss,
tesla, etc.). For instance the value 1.0203 could mean 1.0203 G or
1.0203 T.
:CALCulate#:HOLD:PEAK:STATe <b>
The PEAK HOLD function is turned on when <b> is “1” or “ON”, or turned
off when <b> is “0“ or “OFF”.
:CALCulate#:HOLD:PEAK:STATe?
The state of the PEAK HOLD function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate#:HOLD:PEAK:CLEar
The value being held by the PEAK HOLD function is reset to 0.
:CALCulate#:HOLD:PEAK?
The value being held by the PEAK HOLD function is placed in the output
queue and will be between 0.0000000000 and
±
29999999. The meaning
of this value depends upon the currently selected unit of measure (gauss,
tesla, etc.). For instance the value 1.0203 could mean 1.0203 G or
1.0203 T.
Summary of Contents for 7030
Page 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Page 24: ......
Page 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Page 26: ......
Page 122: ...THIS SIDE BLANK Back B 3 ...
Page 124: ...THIS SIDE BLANK Back of Warranty ...
Page 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Page 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...