IQ-3100 and IQ-3300 Remote Control
Variable Attenuator/Variable Reflector
77
IQ-3100/IQ-3300 Specific Commands
OUTPut:APMode?
Description
This query returns values in terms of the selected display
mode.
Syntax
OUTP(0..26):APM?
Response
“0”-Relative display mode is selected
“1”-Absolute display mode is selected
“2” X+B display mode
Example
OUTP3:APM?
OUTPut[:STATus]
Description
This command controls the shutter. With an attenuator, the
shutter blocks optical continuity. For a variable reflector, the
shutter removes the reflection generating mirror from the
circuit and optical continuity is maintained.
Syntax
OUTP(0..26)[:STAT]<space><boolean>
Parameters
The <boolean> parameter can be
“0” or “OFF”-closes the shutter optically.
“1” or “ON”-opens the shutter optically.
Example
OUTP3:STAT ON