IQ-3100 and IQ-3300 Remote Control
74
IQ-3100/IQ-3300
IQ-3100/IQ-3300 Specific Commands
INPut:REFerence?
Description
This query returns the current reference value.
Syntax
INP(0..26):REF?
Response
The current reference in the “±999.999” format. The units
are dB.
Example
INP3:REF?
INPut:REFerence:DISPlay
Description
This command sets the reference value to the current absolute
attenuation or reflection setting. The displayed attenuation or
reflection will be relative to the reference.
Syntax
INP(0..26):REF:DISP
Example
INP3:REF:DISP
INPut:WAVelength
Description
This command selects a specific calibrated wavelength. The
wavelength range depends on the type and configuration of the
module. See the appropriate module instruction manual for
further information.
Syntax
INP(0..26):WAV<space><numeric_value>[<space><NM>]
Parameters
The <numeric_value> parameter is the current wavelength in
the format 9999.9[NM]. The units are optional.
Example
INP3:WAV 1310 NM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com