5. Installation
DEPOLOX® Pool Compact
38
EN
5.6
Removing and fittting the
housing cover
5.6.1 Flow cell
Removing
•
Remove the housing cover of the flow
cell. To do this, press both unlocking but-
tons (A) on the top of the housing and
carefully remove the cover toward the
front.
Fitting
•
Fit and engage the housing cover of the
flow cell. To do this, position the housing
cover at the bottom and carefully press it
upward until the housing cover engages
on the unlocking buttons.
Fig. 1
View of top of housing cover
A Unlocking buttons
5.6.2 Electronics module
Removing
1 Release the four screws on the housing
cover.
2 Remove the housing cover carefully.
3 Hook the housing cover into the holders
(A) on the basic housing.
Fig. 2
Section, housing cover
A Holders
Fitting
1 Place the housing cover carefully on the
basic housing.
2 Tighten the housing screws by hand (to a
maximum torque of 0.5 ft-lb ± 0.11 ft-lb
(0.7 Nm ± 0.15 Nm)).
5.7
Installing the optional strainer
1 Release the screw joint on the sample
water inlet with ball valve (A) (threaded
connection G 1/2” A).
2 Connect strainer with pipe clamp (B).
3 Connect sample water inlet (C).
Fig. 3
Section, installation of strainer
A Screw joint on sample water inlet with ball
valve
B
Strainer with pipe clamp
C
Sample water inlet
5.8
Connecting the sample water
inlet
•
The sample water inlet must be installed
upstream of the flocculant dosing station.
Otherwise, it could influence the measu-
rements.
•
The sample water inlet must be installed
according to the local regulations in the
pool return or as a pool extraction.
•
The sample water inlet must be chosen to
ensure that the water sample is represen-
tative with a constant, bubble-free flow
and a constant pH value (range from 6 to
8).
•
A distinction must be made between a
sample water inlet with tubing connec-
tion and a sample water inlet with rigid
pipework.
•
PVC hose ø 6x3 or PVC pipe DN 6.
A
A
A
NOTICE
No water pipes made of copper may be fit-
ted in the installation. They would falsify the
measurement.
A
B
C