DEPOLOX® Pool Compact
3. Description
23
EN
3.8.3 Sensors
NOTICE
When disinfecting with inline electrolysis systems or hydrogen drainage into the pool water,
the gold version of the chlorine or ORP (Redox) sensor must be used.
Chlorine sensor
platinum version (W3T160652)
gold version (W3T160991)
Version
Amperometric 3-electrode sen-
sor with platinum electrodes, salt
reserve, zirconium dioxide dia-
phragm, polymerized solid elec-
trolyte, Ag/AgCl reference
electrod
Amperometric 3-electrode sen-
sor with gold electrodes, salt
reserve, zirconium dioxide dia-
phragm, polymerized solid elec-
trolyte, Ag/AgCl reference
electrode
Measurement range
0 bis 50 mg/l Cl
2
Working temperature
range
-5 to +80°C (23 to 176°F)
Operating pressure
0 to 6 bar (6 x 10
5
Pa)
Minimum conducti-
vity of the sample
water
50 µS/cm
Installation length
165 mm
Screw-in thread
PG 13,5
Storage temperature
-5 to +30°C (23 to 86°F)
OPR (Redox) sensor
platinum version (W3T169298)
gold version (W3T172356)
Version
Combination electrode with plati-
num electrode, salt reserve,
zirconium dioxide diaphragm,
polymerized solid electrolyte, Ag/
AgCl reference electrode
Combination electrode with gold
electrode, salt reserve, zirconium
dioxide diaphragm, polymerized
solid electrolyte, Ag/AgCl refe-
rence electrode
Measurement range
±2000 mV
Working tempera-
ture range
23 to 176°F (-5 to +80°C)
Operating pressure
0 to 6 bar (6 x 10
5
Pa)
Minimum conducti-
vity of the sample
water
50 µS/cm
Installation length
120 mm
Screw-in thread
PG 13.5
Storage temperature