152
153
10
ELECTROMAGNETIC IMMUNITY
IEC 60601 test level
Compliance level
Electrostatic discharge (ESD)
IEC 61000‑4‑2
+/‑ 8 kV contact
+/‑2 kV, +/‑4kV, +/‑8kV +/‑15kV air
+/‑ 8 kV contact
+/‑2 kV, +/‑4kV, +/‑8kV +/‑15kV air
High‑frequency electromagnetic
fields
IEC 61000‑4‑3
10 V/m
80 MHz ‑2.7 GHz
80% AM at 1 kHz
10 V/m
80 MHz ‑2.7 GHz
80% AM at 1 kHz
27 V/m
385 MHz PM 18Hz
27 V/m
385 MHz PM 18Hz
28 V/m
450 MHz PM 18Hz
28 V/m
450 MHz PM 18Hz
9 V/m
710 MHz PM 217 Hz
745 MHz PM 217 Hz
780 MHz PM 217 Hz
9 V/m
710 MHz PM 217 Hz
745 MHz PM 217 Hz
780 MHz PM 217 Hz
28 V/m
810 MHz PM 18 Hz
870 MHz PM 18 Hz
930 MHz PM 18 Hz
28 V/m
810 MHz PM 18 Hz
870 MHz PM 18 Hz
930 MHz PM 18 Hz
28 V/m
1720 MHz PM 217 Hz
1845 MHz PM 217 Hz
1970 MHz PM 217 Hz
28 V/m
1720 MHz PM 217 Hz
1845 MHz PM 217 Hz
1970 MHz PM 217 Hz
IEC 60601 test level
Compliance level
Recommended separa‑
tion distance
RF interference
currents
IEC 61000‑4‑6
3 Vrms
150 kHz ‑ 80 MHz
3 Vrms
150 kHz ‑ 80 MHz
𝑑
= 1.2
𝑃
80 MHz - 800 MHz
Radiated RF
disturbances
IEC 61000‑4‑3
10 V/m 80 MHz ‑ 2.7 GHz
10 V/m 80 MHz ‑ 2.7 GHz
𝑑
= 1.2
𝑃
80 MHz - 800 MHz
𝑑
= 2.3
𝑃
800 MHz - 2.5 GHz
Summary of Contents for Pharma D-mine Pump
Page 1: ...EVER Pharma D mine Pump Instructions for use ...
Page 2: ...WELCOME ...