2
Set end
Use
+
and
–
to set the end of the
current basal period and confirm with
.
If the current basal period is
intended to be the last of the day,
then the end time should be set
to match the start time of basal
period 1. Any intervening basal
periods will then be overwritten.
Set basal rate
Use
+
and
–
to set the basal
rate for the current basal period and
confirm with
.
Repeat the two steps for all other
basal periods. For the last basal
period of the day, the end time should
be set to match the start time of basal
period 1 (7:00 AM in the example).
6.1.3 SETTING THE REMAINING BASAL PERIODS
The end of basal period 1 is
automatically the start of basal
period 2. The same applies for all
subsequent basal periods. You
therefore only need to enter the
end time and relevant basal rate
for all subsequent basal periods.
Confirm profile
The pump will display an overview of
all basal periods with your basal rates,
the total of all drug deliveries, and the
delivery time periods. Check all
settings against your table.
If everything is correct, then press
to confirm.
6.1.4 FINALISING THE PROGRAMMING
Once you have covered the whole
day with your entries, a diagram of
the entire daily profile will appear.
The new programming of the pump
will not be applied until you confirm
the entries.
If not, press
to return to pro gramming
or
to discard the changes.
If you discard the displayed
settings, then the programming
will remain unchanged. Return to
"Setting basal period 1" and enter
the new values to make changes.