118
119
7
7.5.2 SIGNAL LIGHT ON THE
DOCKING STATION IS FLASHING
Signal light
Colour
Meaning
What should I do?
Rechargeable
battery for
pump
flashes yellow
The pump is not completely
inserted into the docking station
or the rechargeable battery is
not connected to the pump
or the rechargeable battery is
defective
Check whether the pump has been
placed correctly in the docking
station.
Insert the rechargeable battery into
the pump.
If necessary, replace the rechargeable
battery in the pump as quickly as
possible.
Spare
rechargeable
battery
flashes yellow
Rechargeable battery defective
Replace the rechargeable battery in
the pump as quickly as possible and
replace the defective rechargeable
battery with a new one. Contact
Customer Service.
If one of the signal lights on the docking station is
flashing, one of the following errors has occurred:
This section gives some common examples of when a pump might not function
correctly. We also provide recommendations concerning how to safely continue
operating your pump.
7.6
TROUBLESHOOTING
Please contact Customer Service if the
information in this section is not sufficient
to continue operating your pump safely.