Notice
Why did it occur?
What should I do?
i
Bolus maximum
reached
The "Bolus maximum reached" notice will appear
when the maximum possible number of boluses
to be delivered per day has been reached.
Press
to acknowledge the notice and wait
until the lockout time has elapsed. You can then
deliver another bolus.
i
Bolus button pressed
too long
The "Bolus button pressed too long" notice will
appear when the bolus button is pressed and
held for longer than 30 seconds.
Press
to acknowledge the notice and follow
the instructions for bolus delivery in
S
ection
4.2
.
i
Function not available
Drug delivery is switched on and you attempt to
execute one of the following functions:
– Reset
– Set time
– Prime infusion set
Stop delivery and then execute the desired
function.
Do not forget to then restart drug delivery!
i
Value will increase
The "Value will increase" notice will appear when
the original value is exceeded by at least 100%
when setting a value.
Check to ensure that you have entered the correct
value and confirm the notice with
.
i
Value will significantly
decrease
The "Value will significantly decrease" notice will
appear when the set value falls below the original
value by at least 50%.
Press
to acknowledge the notice and check
to ensure that you have entered the correct value.
Notice
Why did it occur?
What should I do?
i
Vial incorrect
The "Vial incorrect" notice will appear when an
invalid or previously used vial is detected by the
pump.
Press
to acknowledge the notice, dispose
of the vial with the household waste, and fill the
reservoir with a new vial in accordance with the
instructions in
Section
3.1
.
i
No vial
The "No vial" notice will appear when an invalid or
previously used vial is detected by the pump.
Press
to acknowledge the notice and fill the
reservoir with a new vial in accordance with the
instructions in
Section
3.1
.
i
Code incorrect
The "Code incorrect" notice will appear when an
incorrect code has been entered for a password‑
protected pump function.
Press
to acknowledge the notice and enter
the correct code on
page
161
.
i
Reject settings
The "Reject settings" notice will appear when
pump programming has not been confirmed with
.
All changes are discarded. Press
to
acknowledge the notice.
i
Start button
pressed too long
The "Start button pressed too long" notice will
appear when the start button has been pressed
and held for longer than 30 seconds.
Press
to acknowledge the notice and follow
the instructions for starting the pump on
page
57
.