The pump will trigger an alarm if a
user response is required in order to
ensure the continued, safe delivery of
the drug.
The pump will immediately stop drug
delivery in the case of an alarm and
emit an alarm signal to indicate the
interruption.
The tables on the following pages
describe the various alarms and
troubleshooting measures to resolve
each issue.
7.2
ALARMS
Device error
The red signal light above the display
will flash red in the case of all alarms.
The device will emit an acoustic signal.
If the screen remains blank, then a
device error has occurred.
The device error message will
remain displayed for three minutes. For
additional instructions concerning
device errors, see page 109.
Alarm with error message
The pump will display a known error
message with a text message. The
screen will provide the reason for
the interruption and corresponding
instructions to resolve the issue.
The alarm will remain displayed on
the screen until you acknowledge the
message with
.
Alarm
Reason for alarm
What should I do?
Reservoir empty
The amount of drug remaining in the
reservoir is less than the set basal rate or
bolus dose.
Press
to acknowledge the alarm and
change the infusion set and reservoir
in accordance with the instructions in
Section
4.4
.
No reservoir
The reservoir has been removed during
operation.
Press
to acknowledge the alarm and
reinsert the reservoir. Then start the delivery
of the basal rate.
If you wish to connect a new reservoir,
follow the instructions in
Section
4.4
.
Rechargeable
battery empty
The rechargeable battery is empty and
needs to be recharged.
Press
to acknowledge the alarm
and replace the rechargeable battery in
accordance with the instructions in
Section
4.7
.