Description
What should I do?
Pump will not
stop
If the pump will not stop, this may be due to the following:
The "Stop" button was held and pressed too long or not long enough. Follow the corresponding
instructions in
Section
3.4
.
Reservoir is not
fully filled with
liquid
The fit of the adapter was not checked before the filling process. If this is too loose, air can be drawn in
during the filling process. Take a new reservoir and a new vial. Before you start the filling process, check
the placement of the adapter on the reservoir in the pump. To do so, gently turn the adapter with the vial to
the right (clockwise). Repeat the filling process according to the instructions in
Section
3.1
.
Bolus cannot be
delivered
If the bolus cannot be delivered, this may be due to the following:
Drug delivery has stopped and must be restarted. Follow the corresponding instructions in
Section
3.4
.
The bolus lock is currently enabled. It is not possible to deliver a bolus before the lockout time has elapsed.
The available number of delivered boluses has been reached. Another bolus delivery is therefore not possible
until after the lock time has elapsed.
The bolus dose exceeds the amount of drug remaining in the reservoir.
Description
What should I do?
The rechargeable
battery is empty
or not fully
charged
If the rechargeable battery is empty or not fully charged, this may be due to the following:
The rechargeable battery was removed prematurely from the docking station. Make sure not to remove
the rechargeable battery from the docking station until it is completely charged. Follow the corresponding
instructions in
Section
4.7
The lifetime of the rechargeable battery has expired or the docking station is damaged and is no longer
charging the battery. Observe the signal lights on the docking station as described in
Section
1.2
and
notify Customer Service concerning any displayed errors.
New reservoir
not able to be
filled
If the new reservoir is not able to be filled, this may be due to the following:
"Same reservoir" has been incorrectly selected on insertion of a new reservoir. Remove the reservoir and
reinsert, select "New reservoir" and follow the instructions in
Section
3.1
.
Incorrect entry
during insertion
of the reservoir
"Same reservoir" instead of "New reservoir" has been incorrectly selected on insertion of a new reservoir.
Please proceed as follows:
Remove the reservoir and reinsert, make the correct selection and follow the instructions in
Section
3.1
.