74
75
4
3
3
2
2
1
The pump's History function allows
you to review all the important events
of the last three days. The following
information is saved:
•
Filling of the reservoir and the
infusion set
•
Alarms
•
Warnings
•
Boluses
•
Changes of the delivery settings
•
Changes of the time and sound
volume settings
•
Insertion of the rechargeable
battery
Selecting the history function
Select the
History
function in the
menu and press
. You will see an
overview of the last three days with
the total amount (basal and boluses)
and the number of boluses.
Select one of the days and press
.
Press
to return to the main menu.
4.6
VIEWING THE
HISTORY
1
Stopping drug delivery
Stop drug delivery before changing
the rechargeable battery unless delivery
has already stopped automatically,
e.g. due to an alarm.
Section
3.4
Inserting a charged battery
Remove the charged spare battery
from the docking station and insert it
into the pump.
The pump will switch on.
Insert the empty rechargeable
battery in the docking station to
recharge it in order to prevent a
disruption in treatment the next
time the rechargeable battery is
replaced because no charged
battery is available.
Removing an empty battery from
the pump
Press and hold the button on the
rechargeable battery at the back of
the pump and remove the battery
from the battery compartment. The
pump will switch off.
When the battery is removed, all
settings are preserved.
4.7
CHANGING
THE RECHAR‑
GE ABLE BATTERY
Summary of Contents for Pharma D-mine Pump
Page 1: ...EVER Pharma D mine Pump Instructions for use ...
Page 2: ...WELCOME ...