U2829 Series Operation Manual Chapter 5 Execute LCR Operation and Some Examples
61
r)
Insert the short plate (U26010) into the test fixture. Please ensure that the
short plate and the reeds of the test fixture have good contact.
s)
Press the soft key SPOT SHORT to execute short correction.
t)
Insert a standard capacitance into the test fixture. Please ensure that the pins
of the standard capacitance have good connection with the reeds of the test
fixture.
u)
Press the soft key MEAS LOAD to execute load correction.
2)
Note
a)
Because of the different software editions, the soft keys and status
information may be different form this book, but it may not affect users’
understanding.
b)
The load correction is only valid for the components with the same
specification. If the specification is changed, it is required to redo load
correction.