U2829 Series Operation Manual Chapter 3 Introduction to [DISP]
28
Δ
ABS
Δ
%
OFF
6)
Use above soft keys to se the deviation mode of the primary parameter.
7)
Move the cursor to the
DEV B
zone, the following soft keys will be displayed.
Δ
ABS
Δ
%
OFF
8)
Use above soft keys to se the deviation mode of the secondary parameter.
3.5
<CORRECTION>
Press
[SETUP]
to select
CORRECTION
to enter into the
<CORRECTION>
page.
U2829307
Open, short and load correction on the
<CORRECTION>
page can be used to
eliminate the distribution capacitance, spurious impedance and other
measurement errors. U2829 provides two correction modes: the first one is
executing open and short correction on all frequency points through interpolation
method; the other one is executing open, short and load correction on the
frequency point currently set. The following measurement control parameters
can be set on the
<Correction>
page.
Open correction (
OPEN
)
Short correction (
SHORT
)
Load correction (
LOAD
)
Cable length selection (
CABLE
)
Single/ multiple correction mode selection (
CH MODE
)