U2829 Series Operation Manual Chapter 6 Performance and Test
74
Open and short correction should be made before testing. Connect standard AC
resistors: 10
Ω
, 100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
and change the frequency. The error
between reading and nominal value should be in the range ruled in this chapter.
6.6.10
Accuracy of DCR
Test condition
Function DCR
Test frequency -----
Level -----
Range AUTO
Bias -----
Speed Slow
Short correction should be made before testing. Connect standard DC resistors:
0.1
Ω
, 1
Ω
, 10
Ω
, 100
Ω
, 1k
Ω
, 10k
Ω
, 100k
Ω
. The error between reading and
nominal value should be in the range ruled in this chapter.