U2829 Series Operation Manual Chapter 7 Command Reference
94
When the measurement parameter is inductance (Zx), secondary parameter
<DATA B> is the phase angle
θ
, its description is the same as that of <DATA A>.
When the measurement parameter is inductance (Acr), secondary parameter of
<DATA B> is inductance X, its description is the same as that of <DATA A>.
When the measurement parameter is inductance Lx, the secondary parameter of
<DATA B> is the quality factor Q, its description is the same as that of <DATA A>.
Meanwhile when the measurement serial number is equal to or larger than 5 (that
is to say testing leakage inductance LK and DC resistance DCR), no secondary
parameter will be displayed but directly display the status. The status format is :
SN: +/_, N is 0/1.
Status Description
-1 Low
0 Pass
1 High
7.1.14
CORRection subsystem commands
The CORRection subsystem commands are mainly used to set the correction
function, OPEN, SHORT, LOAD.