Output Parameters:
A character string that represents the version of ETSProbe DLL.
ETS_SUPPLYVOLTAGE()
int ETS_SupplyVoltage(int Handle, float &voltage);
Purpose:
Monitors the converter voltage of the laser. In addition the battery voltage
of the probe is accessible through this function.
ETS_ReadBattery
function
must be called to refresh this value.
Return Value:
Returns an integer status code. The numeric value
0
indicates successful
completion. See
Status Codes
for error code description.
Input Parameters:
int probeHandle
Output Parameters:
The converter voltage or battery voltage.
ETS_LASERTEMPERATURE()
int ETS_LaserTemperature(int Handle, float °reesC);
Purpose:
Monitors the laser temperature.
ETS_ReadTemperature
must be called to
refresh this value.
Return Value:
Returns an integer status code. The numeric value
0
indicates successful
completion. See
Status Codes
for error code description.
Input Parameters:
int probeHandle