75
74
Input Parameters:
int probeHandle
Int range:
Accepts values 0–5.
Output Parameters:
None.
ETS_SETUNITS()
int ETS_SetUnits(const int probeHandle, int Unit);
Purpose:
Switches the field units for the probe. The available units are probe-
dependent. Consult the probe user manual for more information.
Return Value:
Returns an integer status code. The numeric value
0
indicates successful
completion. See
Status Codes
for error code description.
Input Parameters:
int probeHandle
Int unit:
Accepts values 1–3.
Output Parameters:
None.
ETS_TEMPERATUREC()
int ETS_TemperatureC(const int probeHandle, int &TempC);
Purpose:
Reads the internal temperature of the probe.
ETS_InitiateReadTemperature
must be called prior to this function call. Use the
ETS_IsOperationComplete
to check for operation completion.
Return Value:
Returns an integer status code. The numeric value
0
indicates successful
completion. See
Status Codes
for error code description.