Charging Indicators
The following LEDs are located on the front of the battery charger:
• POWER ON
(green)—Indicates the battery charger is connected to
the AC power source.
• NO BATTERY
(amber)—Indicates the battery charger does not detect
a battery.
• PENDING
(amber)—Indicates the battery charger detects a battery.
Before fast charging can begin, the battery voltage must fall within
predetermined acceptable limits. A pulse-trickle charge is provided to
bring a depleted battery to a valid charge prior to fast charge.
• CHARGING
(amber)—Indicates the voltage pre-qualification condition
has been met, and fast charge has started. Fast charging continues
until termination by a peak voltage detect or a maximum charge
time. For peak voltage detect, the fast charge is terminated when the
battery voltage is lower than the previously measured values by 24mV.
If a peak voltage is not detected, then the fast charge is terminated
when the maximum time limit of 180 minutes is reached.
• COMPLETE
(green)—Indicates a fast charging peak voltage is
detected. If the probe is turned on during the fast charge period of
the charge cycle, the current surge that the batteries provide to the
electronics in the probe will cause a momentary voltage drop on the
batteries, causing the charger to terminate the fast charge based on a
peak voltage detect condition.
The battery charger switches to top-off charge mode that provides a
charge current one-eighth of the fast charge rate. This charge continues
from the time when a peak voltage detect terminates the fast charge,
until the maximum time of 180 minutes is reached. If the probe is
disconnected from the charger at this point in the charge cycle, the
batteries will operate properly.
Following the maximum time of 180 minutes, a pulse-trickle charge mode
is used to compensate for the self-discharge of the batteries while idle in
the charger.
The fast charge current is low enough that there is not always enough
heating of the battery cells to cause the voltage to drop. With no
voltage drop there is no peak voltage detection, and the fast charge will
terminate after the maximum time of 180 minutes is reached.