24
Battery-Operated Probe Specifications
HI-6023 Specifications
Dynamic Range:
2.0–800 Volts per meter (V/m)
Ranges:
Single
Typical Frequency
Response:
10 kHz–1 GHz
•
10 kHz–30 kHz: +0.5 dB, -2.5 dB
•
30 kHz–1 GHz: ±1 dB
Typical Frequency
Response with Correction:
10 kHz–1 GHz: ±0.9 dB
Linearity:
±0.5 dB
Isotropicity:
±0.5 dB @ 400 MHz
Overload Withstand:
> 1,500 V/m CW
Fiber Optic Cable
Connector:
Standard FSMA
Probe Mount:
1/4–20 UNC tapped hole (internal thread)
Battery:
Rechargeable Nickel-Metal Hydride (NiMH)
Battery Life:
Up to 8 hours
Battery Charger:
•
100–240 VAC universal input
•
3-hour charge from full depletion
Environmental
Operating Temperature:
10°C to 40°C
50°F to 104°F
Humidity:
5% to 95% relative humidity, non-
condensing
Dimensions
Housing:
32 mm x 32 mm x 32 mm
1.26 in x 1.26 in x 1.26 in
Probe Shields:
36 mm (1.42 in)
Weight:
80 g (2.82 oz)