HI-6105 Specifications
Dynamic Range:
0.5–800 Volts per meter (V/m)
Single Range (64 dB)
Typical Frequency
Response:
100 kHz–6 GHz
•
500 kHz–2 GHz: +1 dB, -2.5 dB
•
2 GHz–5.5 GHz: +3.5 dB, -4.5 dB
•
5.5 GHz–6 GHz: +2 dB, -6 dB
Typical Frequency
Response with Correction:
100 kHz–6 GHz: ±0.9 dB
Linearity:
±0.5 dB (1–800 V/m)
Isotropicity:
±0.5 dB @ 400 MHz
Overload Withstand:
1,500 V/m
Physical Interface:
•
Duplex optical fiber (62.5 micron
multimode)
•
FC connectors for laser cable, integral
1-m optical cable
•
ST connector for transmitter cable,
integral 1-m optical cable
Probe Mount:
1/4–20 UNC (internal thread)
Environmental
Operating Temperature:
10°C to 40°C
50°F to 104°F
Humidity:
5% to 95% relative humidity, non-
condensing
Dimensions
Housing:
32 mm x 32 mm x 32 mm
1.26 in x 1.26 in x 1.26 in
Probe Shields:
43 mm (1.69 in)
Weight:
0.08 kg (2.6 oz.)