29
28
Note:
Before using
your battery-operated
probe, read their
respective sections in
the Appendices.
To calibrate the probe
prior to shipment, ETS-
Lindgren also charges
the internal battery
at the factory. Every
effort is made to make
sure that the probe
arrives ready to use, but
you should verify the
condition of the battery
prior to making any
measurements.
For a list and
description of
communication and
information transfer
protocols, including
command structure
and probe commands,
see Appendix D:
Operating Protocols.
BATTERY-OPERATED FIELD PROBES
The HI-60XX Series Field Probe is a battery-operated probe that utilizes three
orthogonal sensors to provide an isotropic reading of the electric field. When
requested, X, Y, Z, and total field data can be reported, making the HI-60XX
Series true 3-axis probes.
The HI-60XX Series uses either the HI-4413P Fiber Optic Modem or the HI-
4413USB Fiber Optic to USB Converter to communicate with ProbeView II™
software through a port on the computer. The HI-6100 Field Monitor may also be
used with the HI-60XX Series.
HI-6023 Field Probe
The
ETS-Lindgren HI-6023 Field
Probe
provides broadband frequency
coverage and wide dynamic range
that satisfies the demands of most
test requirements.
The frequency response of the HI-
6023 is 10 kHz to 1 GHz, and the
dynamic range is 2 to 800 Volts per
meter (V/m).
The HI-6023 is a true 3-axis probe. When requested, X, Y, Z, and total field
data can be reported.
HI-6023 Power Switch
The power switch activates and deactivates the HI-6023:
ON (I)
—When the power switch
is in the I position, an internal 4.8
VDC Nickel-Metal Hydride (NiMH)
battery provides power to the
probe, and the power indicator LED
blinks.
OFF (0)
—When the power switch
is in the
O
position, the probe is
inactive. To prolong battery life, set
the switch to
OFF
when the probe
is not in use.