54
Note:
The BP
command is provided
for backward
compatibility and
should not be used to
monitor the converter
voltage, which
always responds with
:B64N<CR>.
Note:
The Identification
command, I, may
also be used as the
first command sent.
The command will
turn on the laser.
Once communication
between the probe and
the Laser Data Interface
(LDI) is established,
the return string will
be sent. Subsequent
I commands
immediately send the
return string.
Probe Commands
All probe commands return :E7 when the probe is turned off.
HI-61XX Series Laserpro Field Probes
Probe Command
Description
Probe Response
B<CR>
Read probe converter voltage
:Bxx.xx<CR>
BP
Read probe converter voltage in
hexadecimal format
:B64N
•
N=safe operating level
•
F=fail level
Voltage reported as 0–64;
64 corresponds to 100%
D3
Read probe data
:Dx.xxxyy.yyzzz.zB<CR>
•
xxxx, yyyy, zzzz=
4-digit axis values with
floating decimal point
•
B=battery flag, N or F
D5
Read probe data
:Dx.xxxyy.yyzzz.zcccc.B<CR>
•
xxxx, yyyy, zzzz=
4-digit axis values with
floating decimal point)
•
cccc=composite field
value with floating
decimal point
•
B=battery flag, N or F
I
Identification command
:I61XX<sr><sn><cd>B<CR>
•
XX=specific probe model
•
sr=10-character software
revision
•
sn=8-character serial
number
•
cd=8-character
calibration date
•
B=battery flag, N or F