Figure 3m: Programmable Bake Flow Control
Once the desired options are selected, <click> the
Exit
button to return to the
method parameters screen.
Bake Time (min) –
This is the total length of time for one bake cycle. The bake time can
programmed from 0.0 to 999.9 minutes.
Bake Cycles
– The number of bake cycles for one sample analysis. This can be
programmed from 1 to 10 cycles. After the first bake cycle is completed the trap and
MoRT will cool to approximately 60°C before beginning to heat for the next bake cycle.
Printing the Method
-
<click> on the Printer icon. The method can be printed if
the Encon EV is interfaced to a printer
Saving the Method
-
<click> on the
Save
icon and the method will be saved.
NOTE: After saving a method, it must be opened to activate.
Open a Saved Method -
<
click> on the
Open
icon and all previously saved
Methods are available to be loaded.
Summary of Contents for Encon Evolution
Page 1: ...Encon Evolution User Manual Original Operating Instructions Rev 4 5 ...
Page 2: ......
Page 8: ...User Manual Introduction Chapter 1 ...
Page 15: ...User Manual Installation and Setup Chapter 2 ...
Page 20: ...User Manual Operation Chapter 3 ...
Page 47: ...User Manual Maintenance and Troubleshooting Chapter 4 ...
Page 61: ...User Manual Flow Diagrams Chapter 5 ...
Page 62: ...Figure 5 1 Standby Dry Purge Flow Diagram ...
Page 63: ...Figure 5 2 Purge Flow Diagram ...
Page 64: ...Figure 5 3 Desorb Ready Flow Diagram ...
Page 65: ...Figure 5 4 Desorb Flow Control Flow Diagram ...
Page 66: ...Figure 5 5 Desorb Auto Drain Flow Diagram ...
Page 67: ...Figure 5 6 Desorb Flow Diagram ...
Page 68: ...Figure 5 7 Bake with Bake Gas Bypass Flow On Flow Diagram ...
Page 69: ...Figure 5 8 Bake with Bake Gas Bypass Flow Off Flow Diagram ...
Page 70: ...Figure 5 9 Leak Check Flow Diagram 1 ...
Page 71: ...Figure 5 10 Leak Check Flow Diagram 2 ...
Page 72: ......