![EST Encon Evolution User Manual Download Page 38](http://html.mh-extra.com/html/est/encon-evolution/encon-evolution_user-manual_2429115038.webp)
Bake Flow Rate (ml/min) –
This is the flow rate of the pressurization gas during the
bake cycle. The bake flow rate can be programmed from 0 to 500ml/min in 1ml/min
increments.
NOTE: It is suggested to have a minimum a flow rate of at least 10ml/min
passing through the trap when the trap temperature is above ambient. This will
prevent damage to the trap material.
Programmable Bake Flow Control -
<click> the
Program
button
to enable this feature.
The screen as shown in Figure 3m will appear. <Check > the
Enable Ramp Control
box to program the desired settings.
Initial Flow (ml/min) –
This the beginning bake flow rate programmed in
ml/min. The initial flow can be programmed from 0 to 500ml/min in 1ml/min
increments.
Initial Hold Time (min) –
This is the time in minutes the initial flow is
maintained. The initial hold time can be programmed from 0.0 to 999.9 minutes.
Ramp Rate ((ml/min)/min) –
This is the rate programmed in ml/min the bake
flow will increase after the initial hold time has expired to the desired final flow
set point. The ramp rate can be programmed from 1 ml/min per minute to 500
ml/min 1 ml/min increments.
Final Flow (ml/min) –
This is the ending bake flow rate programmed in ml/min.
The final flow can be programmed from 0 to 500ml/min in 1ml/min increments.
Final Hold Time (min) –
The final hold time is not a value to be entered by the
User. The final hold time is the calculated difference of the final bake time taken
from the sum of the initial time and the ramping time needed to achieve the final
flow rate.
For Example: Total Bake Time = 10 minutes
Initial Flow = 40ml/min
Initial
Hold
Time
=
2
minutes
Ramp
Rate
=
40ml/min
per
minute
Final
Flow
=
240ml/min
Final
Hold
time
=
3
minutes.
Explanation: Since it takes 5 minutes to ramp from 40ml/min to
240ml/min at a rate of 40ml/min per minute. The final hold time is
calculated to be 3 minutes.
10 minutes – (2 m 5 minutes) = 3 minutes
Summary of Contents for Encon Evolution
Page 1: ...Encon Evolution User Manual Original Operating Instructions Rev 4 5 ...
Page 2: ......
Page 8: ...User Manual Introduction Chapter 1 ...
Page 15: ...User Manual Installation and Setup Chapter 2 ...
Page 20: ...User Manual Operation Chapter 3 ...
Page 47: ...User Manual Maintenance and Troubleshooting Chapter 4 ...
Page 61: ...User Manual Flow Diagrams Chapter 5 ...
Page 62: ...Figure 5 1 Standby Dry Purge Flow Diagram ...
Page 63: ...Figure 5 2 Purge Flow Diagram ...
Page 64: ...Figure 5 3 Desorb Ready Flow Diagram ...
Page 65: ...Figure 5 4 Desorb Flow Control Flow Diagram ...
Page 66: ...Figure 5 5 Desorb Auto Drain Flow Diagram ...
Page 67: ...Figure 5 6 Desorb Flow Diagram ...
Page 68: ...Figure 5 7 Bake with Bake Gas Bypass Flow On Flow Diagram ...
Page 69: ...Figure 5 8 Bake with Bake Gas Bypass Flow Off Flow Diagram ...
Page 70: ...Figure 5 9 Leak Check Flow Diagram 1 ...
Page 71: ...Figure 5 10 Leak Check Flow Diagram 2 ...
Page 72: ......