![EST Encon Evolution User Manual Download Page 16](http://html.mh-extra.com/html/est/encon-evolution/encon-evolution_user-manual_2429115016.webp)
2.1 General
Information
The system setup and installation is intended to be as thorough as possible. If in doubt at
any point during the installation, please contact EST Analytical toll free at
(800) 283-
3510
, or outside the United States at
(513) 642-0100
. If installation is not directed or
performed by EST Analytical personnel, the operator must be thoroughly familiar with
setup and installation sections of this manual.
NOTE: Failure to follow the installation instructions as set forth could result in
damage to the product and nullify the warranty.
2.2
Work Space Requirements
Dimensions: 17”H x 11”W x 16”D (43.2cm x 27.94cm x 40.6cm)
.
NOTE: Allow for added space on all sides of the instrument for proper ventilation.
The Encon Evolution weighs approximately 45 lbs. (20.4 kg).
2.3 Power
Requirements
Voltage:
115/230 VAC (±10%)
50/60 Hz
10/6.3 Amps
15/20A circuit breaker
Summary of Contents for Encon Evolution
Page 1: ...Encon Evolution User Manual Original Operating Instructions Rev 4 5 ...
Page 2: ......
Page 8: ...User Manual Introduction Chapter 1 ...
Page 15: ...User Manual Installation and Setup Chapter 2 ...
Page 20: ...User Manual Operation Chapter 3 ...
Page 47: ...User Manual Maintenance and Troubleshooting Chapter 4 ...
Page 61: ...User Manual Flow Diagrams Chapter 5 ...
Page 62: ...Figure 5 1 Standby Dry Purge Flow Diagram ...
Page 63: ...Figure 5 2 Purge Flow Diagram ...
Page 64: ...Figure 5 3 Desorb Ready Flow Diagram ...
Page 65: ...Figure 5 4 Desorb Flow Control Flow Diagram ...
Page 66: ...Figure 5 5 Desorb Auto Drain Flow Diagram ...
Page 67: ...Figure 5 6 Desorb Flow Diagram ...
Page 68: ...Figure 5 7 Bake with Bake Gas Bypass Flow On Flow Diagram ...
Page 69: ...Figure 5 8 Bake with Bake Gas Bypass Flow Off Flow Diagram ...
Page 70: ...Figure 5 9 Leak Check Flow Diagram 1 ...
Page 71: ...Figure 5 10 Leak Check Flow Diagram 2 ...
Page 72: ......