![EST Encon Evolution User Manual Download Page 33](http://html.mh-extra.com/html/est/encon-evolution/encon-evolution_user-manual_2429115033.webp)
Final Hold Time (min) –
This is not a programmable setting but a function of the
set conditions. The final hold time will be the difference of the final purge time
from the sum of the initial time and the time needed to achieve the final flow rate.
For Example: Total Purge Time = 4 minutes
Initial Flow = 40ml/min
Initial
Hold
Time
=
0.5
minutes
Ramp
Rate
=
40ml/min
per
minute
Final
Flow
=
200ml/min
The Final Hold time at 200ml/min would be 0.5 minutes.
Explanation: It takes 4 minutes to ramp from 40ml/min to 200ml/min at a
rate of 40ml/min per minute. The final hold time is calculated to be 0.5
minutes.
4 minutes – (0.5 m 3 minutes) = 0.5 minutes
Purge Time (min) –
This is the length of time for the sample purge. The purge time can
programmed from 0.0 to 999.9 minutes.
Trap Dry Purge Temperature (
°
C) –
This is the temperature of the adsorbent trap
during dry purge. The temperature can be programmed from 25°C to the set point in 1°C
increments.
NOTE:
This temperature should be set to below ambient to eliminate the possibility
of analyte migration within the trap prior to desorb.
Dry Purge Flow Rate (ml/min) –
This is the rate the pressurization gas flows through
the adsorbent trap during the dry purge process. The Dry Purge Flow Rate can be
programmed from 1 to 500ml/min in 1ml/min increments.
Programmable Dry Purge Flow Control -
<click> the
Program
button
to enable this
feature. This feature can be used to change dry purge flow rates during the dry purge
cycle. The screen as shown in Figure 3j will appear. Check the
Enable Ramp Control
box to program the desired settings.
Initial Flow (ml/min) –
This is the beginning dry purge flow rate programmed in
ml/min. The initial flow can be programmed from 0 to 500ml/min in 1ml/min
increments.
Initial Hold Time (min) –
This is the time in minutes the initial flow is
maintained. The initial hold time can be programmed from 0.0 to 999.9 minutes.
Ramp Rate ((ml/min)/min) –
This is the rate programmed in ml/min per minute
the dry purge flow will increase after the initial hold time has expired to the
desired final flow set point. The ramp rate can be programmed from 1ml/min to
500ml/min in 1ml/min per minute increments.
Summary of Contents for Encon Evolution
Page 1: ...Encon Evolution User Manual Original Operating Instructions Rev 4 5 ...
Page 2: ......
Page 8: ...User Manual Introduction Chapter 1 ...
Page 15: ...User Manual Installation and Setup Chapter 2 ...
Page 20: ...User Manual Operation Chapter 3 ...
Page 47: ...User Manual Maintenance and Troubleshooting Chapter 4 ...
Page 61: ...User Manual Flow Diagrams Chapter 5 ...
Page 62: ...Figure 5 1 Standby Dry Purge Flow Diagram ...
Page 63: ...Figure 5 2 Purge Flow Diagram ...
Page 64: ...Figure 5 3 Desorb Ready Flow Diagram ...
Page 65: ...Figure 5 4 Desorb Flow Control Flow Diagram ...
Page 66: ...Figure 5 5 Desorb Auto Drain Flow Diagram ...
Page 67: ...Figure 5 6 Desorb Flow Diagram ...
Page 68: ...Figure 5 7 Bake with Bake Gas Bypass Flow On Flow Diagram ...
Page 69: ...Figure 5 8 Bake with Bake Gas Bypass Flow Off Flow Diagram ...
Page 70: ...Figure 5 9 Leak Check Flow Diagram 1 ...
Page 71: ...Figure 5 10 Leak Check Flow Diagram 2 ...
Page 72: ......