4.2.
Removal of LCD assy
1) Remove 2 harnesses connected to the control board Assy.
from the LCD Assy. (CN4 and 5: flat cables): Fig. 1
* CN5 has the lock function of the harness. Remove the
harness from the control board Assy. by unlocking it.
[Fig.1]
Locked
Unlocked
* Move the lock units on both right
and left sides at the same time at
the time of unlocking them.
[Fig.2]
Unscrew the screw (cross-recessed pan head screw M3×8)
fixing the GND harness located behind the lens holder by
lowering the holder.: Fig. 2
Remove the clip fixing the harness from the LCD Assy.: Fig. 3
2) Remove the LCD Assy. by unscrewing the screws (hexagon
socket cap low head bolt M3×6 black) fixing the LCD Assy.:
Fig. 3
[Fig.3]
4.3.
Removal of top cover
Remove the top cover by unscrewing 2 screws
(cross-recessed bind screw M3×8) fixing the top cover.: Fig. 4
[Fig.4]
4.4.
Removal of light receiving unit cover
Remove the lens plate by unscrewing 2 screws (cross-recessed bind screw M3×8 black) on both sides of
right and left.: Fig. 5
Remove the lens holder. The light receiving unit cover is fixed with the click. It can be removed by
lifting A part of the cover.: Fig. 6
[Fig.5]
[Fig.6]
LCD assy.
Control board assy.
Fixing screw : 4
Clip: 3
Top cover
Lens holder
Lens plate
Lens stand
A
Click : right and left
Light receiving unit
cover