8-
Completion of calibration
Calibration is completed.
Now proceed to “28. Finished” by pressing
. Switch to the menu screen by pressing .
The calibration value is stored by switching to the menu screen.
9-
Confirmation of calibration value
Check if calibration is performed correctly by using the test lens used for the calibration.
Open the MENU screen by pressing
on the
measurement screen. Select “1. Setting” on the MENU
screen and display the Setup screen by pressing
.
Select “Step” on the Setup screen with
and set as 0.01 with
Switch to the measurement screen after completing the
setup by pressing
2 times continusouly.
Take a measurement of the lens used for calibration.
(*Perform alignment correctly.)
Ensure that the measurement result is within ±0.06 of
the test value.
31
Marking ok
Perform calibration again if the measurement result of each lens is not within ±0.06.