8.
Calibration
8.1.
How to enter dealer mode
Use the dedicated maintenance lens set when performing calibration.
Perform the calibration in the dealer mode.
In case of ALM700
Turn on the device and hold the Memory/Add switch after the pilot lamp on the bottom right is on (right before
the logo is displayed) until the comment as shown below appears.
8.2.
Device setting
Check the setting of function.
Menu screen is displayed when entering the dealer mode.
Select “3. Function” with
and check the
setting by displaying the setting screen with
.
Setting screen
The setting should be as shown below.
Item
Description of function
Dealer Mode
Mode setting at the time of turning
power on
On: always dealer mode
Off: always normal mode
Error Check
On/ Off of measurement error setting
Adjust
On/ Off of measurement correction
value
Gain Control
On/ Off of auto adjustment of
brightness
of measurement light