The measurement value of the base direction is loaded:
0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°
4-
Calibration of Axis
Input the test value with
.
Place and align the cylindrical lens of +5D so that AXIS
value becomes 90°.
When the prisms reach between 0.00 and 0.01 in both X
and Y directions, CYL and AXIS are automatically
calibrated to the value of the test lens.
Input the test value with
.
Place and align the cylindrical lens of +5D so that AXIS
value becomes 0°.
When the prisms reach between 0.00 and 0.01 in both X
and Y directions, CYL and AXIS are automatically
calibrated to the correction value input.
5-
Calibration of prism
Calibration of 2
△
and 5
△
(
※
2
△
is not included in the maintenance lens set.)
Input the test values with
.
Align the prism lens so that the base direction becomes 0°.
The prism value is automatically calibrated to the test
value.
Adjustment of 10
△
(
※
10
△
is not included in a maintenance lens set.) Input
the test value with
.
Align the calibration lens according to the indication of the
base direction (align B).
When the measurement value of the indicated base
direction is loaded, a beep sounds and the indication of the
base direction changes.
When the measurement value of the last alignment (base
direction: 315°) is loaded, a beep sounds and calibration is
automatically done.
Prism value
Marking ok
Marking ok
Base direction
Indication of base
Direction for alignment
Marking ok