8.3.
How to enter calibration screen
Open the MENU screen by pressing
on the
measurement screen.
Select “5. Calibration” with
and display
the calibration screen with
.
8.4.
Calibration procedure
List of calibration parameter
No.
Item
Setting value
Content
ALM700 ALM500
00
00
Start
Start
03
01
0D
0.00D
OD calibration
04
02
S –5D
-
5.00D
Calibration of spherical values at each setting value (auto setting)
05
03
S –10D
-
10.00D
06
04
S –15D
-
15.00D
07
05
S –20D
-
20.00D
08
06
S –25D
-
25.00D
09
07
S +5D
+
5.00D
10
08
S +10D
+
10.00D
11
09
S +15D
+
15.00D
12
10
S +20D
+
20.00D
13
11
S +25D
+
25.00D
14
12
C +5D A90
+
5.00D
Overall calibration of spherical and cylindrical values and calibration of
angle to the lens plate.
15
13
C +5D A 0
+
5.00D
16
14
P 2
+
2.00
△
Calibration of 2 prism powers
17
15
P 5
+
5.00
△
Calibration of 5 prism powers
18
16
P 10
+
10.00
△
Calibration of 10 prism powers
19
17
Far pt.
0.30 (standard value)
Progressive lens: calibration value of far point 1
20
18
Near pt.
0.38 (standard value)
Progressive lens: calibration value of near point 1
21
19
AT judge
0.42 (standard value)
Progressive lens judgment calibration value 1
22
20
InFa pt
0.12 (standard value)
Progressive lens: calibration value of far point 2
23
21
InNr pt
0.15 (standard value)
Progressive lens: calibration value of near point 2
24
22
InAT jud
0.18 (standard value)
Progressive lens judgment calibration value 2
25
/
PD 0mm
0.00
Calibration value of optical center of PD value (only for ALM700)
26
/
PD
22.5
Calibration value of optical end of PD value (only for ALM700)
28
23
Finished
End
The factory setting value is valid for the items in
. You may not be able to take a measurement
accurately if changing them.
On “5. Calibration”, each setting and loading of the adjustment data are automatically done in order to facilitate
the calibration procedure.