User Manual Rev. 3.0
46
It is essential to perform the inspection and servicing at the intervals indicated
in the MAINTENANCE manual. Failure to do so can have serious adverse outcome,
causing the unit to stop performing as expected and causing damage to samples, pa-
tients or users.
Warranty is considered to be void if service and maintenance procedures are not
followed.
Warranty is considered to be void if service and maintenance procedures are
done not by a trained and authorized personnel.
24 Emergency procedures
Total loss of power to or inside the unit:
•
Remove all the samples and place them in an alternative or backup device that is not
affected by the problem.
•
MAW, Multi-zone ART Workstation
®
will lose its temperature below the safe level. The
speed of temperature drop will depend on the environment temperature.
If a single temperature alarm goes off on the unit:
•
Remove all the samples from affected zone and place them on zones that are not af-
fected by the problem.
If a CO
2
pressure alarm goes off on the unit:
•
Inspect the external gas supply and gas supply lines. If the problem is external and
not readily amendable, normally it will not be necessary to perform any emergency
procedures other than stop using the humidification system.