12.1 MAW
®
without compartments
Figure 12.6
Gas flows through the nozzle
The gas will flow through the nozzle to the working area on a tabletop.
Figure 12.7
Gas nozzle on a tabletop
A gas hood must be placed over the outlet. The constant flow will flush the environment so
that a correct CO
2
concentration could be kept and no pH drift would occur.