MAW
®
User Manual Rev. 3.0
45
Figure 22.10
The report mode
3 action buttons are located at the right-hand side. The RECONNECT button allows the sys-
tem to reconnect with sensors (in case of data loss due to USB connection issues). The button
MINIMIZES toggles the full screen format off in the workstation logger. The button EXIT
shuts down the workstation logger software.
No data will be stored and all safety surveillance functions will be operative if the
data logger is shut down.
The AIO PC can be easily used to display images from any type of USB microscope camera.
A standard USB socket is located under the humidification system. This may be used to
load microscope camera drivers or other software onto the AIO PC. When the software is
set up, the microscope camera can be connected to the USB socket and the image will be
shown on the screen.
23 Maintenance
MAW, Multi-zone ART Workstation
®
is designed to be user-friendly. Reliable and safe oper-
ation of this equipment is based on the following conditions:
1.
Correct calibration of temperature and gas levels, using high-precision equipment in
the intervals that are prescribed on the basis of clinical practice at the laboratory,
where MAW
®
is used. Manufacturer recommends that the period between valida-
tions should be no longer than 14 days.
2.
Replacement of in-line HEPA filter in the correct intervals
–
once per year.
3.
Suitable cleaning, according to the intervals, prescribed by the clinical practice in the
laboratory, where MAW
®
is used. Periods longer than 14 days between cleanings are
not recommended by the manufacturer.